Cold flow properties, such as Cloud Point and Pour Point, indicate how petroleum products behave at low temperatures. The test specimen is cooled at a specified rate and measured periodically. The highest temperature at which a cloud is first detected at the bottom of the test jar is recorded as the cloud point. The lowest temperature at which movement of the specimen is detected is recorded as the pour point.
Meets the requirements of: ASTM D5771, D5950
Excellent Correlation to: ASTM D97, D2500, D5853, D6074, D6158 ; ISO 3015, 3016; IP 15, 219; DIN 51597; FTM 791-201; NF T 60-105; JIS K2269
Detection:
Cloud Point: Optical Detection
Pour Point: Automatic Tilting
Cooling: Internal Cooling System
Temperature Range: -105 °C to +50 °C (-157 °F to +122 °F)
Temperature Accuracy: ± 0.1 °C
Tilt Interval: 1 °C or 3 °C as per test method (User defined for custom method 1 °C to 5 °C)
Interfaces: USB (4), Ethernet
Display:10.4″ Color Touch Screen
Password Protection: Multi-level password capability
- Automatic determination of cloud point by Optical Detection
- Automatic determination of Pour Point via the Automatic Tilt Method
- Meets the requirements of ASTM D5771 and D5950
- Excellent correlation to ASTM D2500, D97, and related specifications
- Wireless Cloud Point Head
- Wireless Pour Point Head
- Integrated Cooling System
- Automatic Calibration by means of 10 point RTD Calibration, Sample Calibration, and Jacket Calibration
- 10.4-inch Integrated Color Touch Screen
-
- Standard test methods are pre-programmed or user-defined test programs can be created
- Clear graph of temperature vs. time for sample and bath temperatures
- Results can be easily searched, viewed on screen, printed, and exported to LIMS
- 4 USB interfaces
- Internet (Ethernet) Line
- 3 Point Electronics Calibration
- 10 Point RTD Offset Calibration
- Jacket Calibration